A capillary column in gas chromatography (GC) is a type of chromatographic column used to separate volatile compounds in a sample based on their interaction with a stationary phase. Capillary columns are distinguished by their small internal diameter (typically between 0.1 and 0.53 mm) and the thin coating of stationary phase on the inside wall. This design leads to high separation efficiency and faster analysis times compared to packed columns.
Structure and Components
Column Material:
Fused silica: Most capillary columns are made of fused silica, a highly pure form of silicon dioxide. Fused silica is flexible, inert, and has excellent thermal stability.
Polyimide coating: The outer surface of the fused silica is coated with polyimide for protection, giving the column flexibility and mechanical strength.
Stationary Phase:
A thin layer of liquid or solid stationary phase is bonded or coated onto the inner walls of the capillary. The composition of the stationary phase determines the types of chemical interactions with the analytes.
Common stationary phases include non-polar (e.g., PDMS – polydimethylsiloxane) for separating compounds based on volatility, and polar (e.g., polyethylene glycol) for separating compounds based on polarity.
Internal Diameter and Length
The internal diameter (ID) of capillary columns typically ranges from 0.1 mm to 0.53 mm. Narrower columns (e.g., 0.18 mm) offer better resolution but require higher pressures.
Column lengths can range from 15 meters to 100 meters. Shorter columns provide faster analysis, while longer columns offer better separation for complex mixtures.
Working Principle
Capillary column gas chromatography operates by injecting a sample into a heated injector, where it vaporizes. The vaporized sample is carried by an inert gas (e.g., helium or nitrogen) through the column. As the sample travels through the column:
Components interact with the stationary phase based on their chemical properties.
Volatile components with lower boiling points move through the column faster, while less volatile components or those with stronger interactions with the stationary phase are retained longer.
The time taken for a compound to travel through the column is called retention time, and it depends on the compound's affinity for the stationary phase and its volatility.
Types of Capillary Columns
Wall-coated Open Tubular (WCOT):
The stationary phase is a thin film of liquid directly coated onto the inner wall of the column.
Support-coated Open Tubular (SCOT):
A solid support, such as diatomaceous earth, is first coated onto the wall, followed by a layer of stationary phase. These columns have more surface area than WCOT, but their performance is slightly lower.
Porous-layer Open Tubular (PLOT):
These columns contain a solid stationary phase layer that can separate gases and volatile compounds more effectively. They are commonly used for gas separations.